Implant Disassembly Due to Pedicle Screw Nut Loosening
Prasad Krishnan, MS, MCh Department of Neurosurgery, National Neurosciences Centre 2nd Floor, Peerless Hospital Campus, 360, Panchasayar, Garia 700094, Kolkata, West Bengal India prasad.krishnan@rediffmail.com
This article was originally published by Thieme Medical and Scientific Publishers Pvt. Ltd. and was migrated to Scientific Scholar after the change of Publisher.
Two male patients (45 years and 58 years old, respectively) underwent spinal decompression, intertransverse grafting and L4–5 fixation with pedicle screw and rod construct for L4–5 grade 1 spondylolisthesis with an uneventful postoperative outcome. Immediate postoperative X-rays and those at 3 months were satisfactory. Follow-up imaging at 6 months in the first case (Fig. 1) showed that the inner nut of the left L4 screw had loosened off and moved out of the construct and was lying just medial to the pedicle screw head. However, the rod remained in position inside the screw head. In the second case, X-rays at 6 months (Fig. 2) revealed that the right-sided rod had migrated medially out off the right L5 pedicle screw head, though in this case the inner nut was still in position. Both were offered resurgery to revise the implant but only the second patient agreed for the same.
Implant disassembly due to pedicle screw nut loosening is a rare complication of spinal instrumentation that predisposes to rod migration with sparse literature mostly in the form of case reports.1 2 3 The inner screw itself, after progressive loosening, may come out of the construct (as in our first case) or may continue to remain in situ (as in our second case). Causes described for screw nut loosening include both insufficient tightening of the inner nut into the pedicle screw head1 3 and improper coupling of the inner nut with the pedicle screw head by not squaring the corresponding threads of both the elements prior to tightening.1 3 The surgeon must be careful and ensure that there is no soft tissue inside the pedicle screw head at the time of tightening the inner nut and that all nuts are completely tightened after the construct is assembled and also take care that, at the time of insertion, the threads of the inner nut match the grooves on the pedicle screw, a process easily achieved by initially rotating the nut in a counter-clockwise direction till a “click” sound is heard.
Conflict of Interest
None declared.
References
- Pedicle screw nut loosening: potentially avoidable causes of spine instrumentation failure. Asian Spine J. 2014;8(2):224-226.
- [Google Scholar]
- Early loosening of spinal rod in a case of degenerative grade 1 spondylolisthesis treated with unilateral pedicle screw fixation and transforaminal cage for interbody fusion. J Orthop Case Rep. 2017;7(4):6-9.
- [Google Scholar]
- Improper coupling between inner nut and screw head leading to rod loosening and dislodgement. Romanian Neurosurgery. 2014;21:245-246.
- [Google Scholar]